10 4 Pre-Installation Requirements 4.1 Tools and Consumable MaterialTable 4-1 provides a consolidated list of the tools and consumable material whic
114 Pre-Installation Requirements Make sure the location allows sufficient space for movement of the arms in the extended condition.4.3.2 Site Envi
124 Pre-Installation Requirements4.3.3 Electrical Outlets & Requirements The mains outlet should have a good Ground connection. Grounding of th
134 Pre-Installation Requirements 4.7 Included System ComponentsFigure 7 shows the major components and accessories included with a wall mounted Int
145 Installation of Intraskan DC 5 Installation Procedure SummaryThis section provides the instructions necessary to install the Wall Mounted Intrask
155 Installation of Intraskan DCFigure 11: Fix base unit rear cover along with power board.• Fix one M10X100 wood screw with M10 plane washer at the
165 Installation of Intraskan DC• Fix the base unit rear cover along with power board assembly to the base unit plate with six M3x6 HSHC screws usin
175 Installation of Intraskan dc 5.2 Procedure for Input wiring:• Strip the insulation on each input wire for a length of approximately 5 mm. Using
185 Installation of intraskan DC 5.3 Straight- Arm Installation:• Remove the circlip fixed to the straight-arm guide rod using circlip pliers as s
195 Installation of intraskan dc 5.4 Scissor Arm Installation • Remove the end cap by removing M3x6 self tapping CSK screws screw using screw drive
2This page is left blankIntentionally.
205 Installation of Intraskan dc 5.5 Scissor Arm Cables ConnectionLay cables flushed along the right inner side. Carefully pull any excess length an
215 Installation of Intraskan dc 5.6 Ground Connection Check:Using a Multimeter, check the continuity between the following points:-• Between Ground
225 Installation of Intraskan dcFigure 32: Wire routing hole & 2 mounting holesFigure 30-: 4 M3x6Hex socket head cap screwsFigure 31: Console cabl
235 Installation of intraskan dc 5.8 IntraskanDCConsolecongurations:Conguration-1:Can use Internal Console and Internal Exposure Switch.Congura
245 Installation of intraskan dcConguration-3:8P8C with door bell switch.Can use two Consoles with Internal Exposure Switch and double door bell swit
255 Installation of intraskan dcConguration-4:3 WIRE WITH SINGLE DOORBELL SWITCH.J1, J2 and J3 Open on External console PCB.Can use Internal Console
265 Installation of intraskan dcConguration-5:3 WIRE WITH DOUBLE DOORBELL SWITCHCan use Internal Console with Internal Exposure Switch and double doo
275 Installation of intraskan dcConguration-66P4C WITH SINGLE DOORBELL SWITCH.J1, J2 and J3 Open on External console extension PCB.Can use Internal C
285 Installation of intraskan dcConguration-76P4C WITH DOUBLE DOORBELL SWITCH.Can use Internal Console with Internal Exposure Switch and double door
295 Installation of intraskan dc 5.9 Baseunitcoverxing:• Take the base unit front cover and connect the console cable to the J5 connector on the
3Table of contents 1 Introduction ...
305 Installation of intraskan dc 5.10 Scissor-Arm Operation CheckingMove the Scissor arm and ensure there is no drift as per steps given below.• Ke
315 Installation of intraskan dcNote: For changing kV follow the steps given below:Restart the unit and while starting press the UP/Increment and DOWN
325 Installation of intraskan dcStep 3:- • Keep the vertical arm in horizontal position and horizontal arm in vertical position as shown in the Fig-u
336 Console Interface 6 Console Interface 6.1 Console as a ToolThe control console of Intraskan DC stores information about the exposures delivered
346 Console InterfaceTable 6-4: Keyboard event flag bitmaskBit Position Description0 Exposure key stuck1~12 <unused>13 Set / Select / Mode key
35 7 Trouble shooting Techniques 7.1 Errors and WarningsWhen in a fault state, the unit would display an error message with a corresponding error co
367 Trouble ShootingTable 7-3: TROUBLE SHOOTING CHART (MECHANICAL)OBSERVED PROBLEM RECOMMENDED ACTION1. Scissor Arm is Drifting from its released posi
37 8 Service Procedure 8.1 Replacement GuideTable 8-1: Replacement Parts listPart Failed Part to be replacedIf the Scissor Arm or parts of Scissor a
388 Service Procedure• Remove the 2 M3X16 self tapping pan head screws using phillips screw driver and remove the bottom cover as shown in the Figure
398. Service Procedure8.2.1.5 HV Tank removal • To remove the HV Tank from the clamp, remove the 5 (M3X6 Hex. socket button head) screws (from 1 to
4Table of Contents 8.2 Replacing the Scissor-Arm: ...
408. Service Procedure8.2.1.6 Final Steps before Removal• Remove 2 bottom caps of the straight arm by removing M3x6 self tapping screws (2 screws fo
418 Service Procedure8.2.2 Installing the New Scissor Arm8.2.2.1 Initial setup• Using gloves, apply grease to the new scissor-arm insertion rod and
428 Service Procedure8.2.2.3 Scissor-arm cable connection: • Using a 2.5mm allen key connect the GND wires(Ring terminals) of both the Communication
438 Service ProcedureFigure 62 • Route the INV power cable as shown in the Figure 62 and connect it to the sealing board 3 pin connector (Pin 1 &
448 Service Procedure• Fix the rubber fixing ring on the positioning cone as shown in Figure 65 (1) and slide the rubber dial ring and fix it in the
458. Service Procedure 8.3 Replacing the Straight-Arm(Applicable for all length)Inverter cable(J1)Communica-tion cable(J2)Ground pointCable tiesFigu
468. Service Procedure• Remove the circlip of the straight arm as shown in Figure 73 and remove the straight arm from the base unit as shown in Figu
478. Service Procedure• Route the scissor-arm cables through the straight-arm till it comes inside the base unit (Use guide wire for routing and guid
488. Service Procedure 8.4 Replacing the Power Board8.4.1 Removing the Power Board• Execute the Steps in “8.2.1.1 Power Off:” above.• Execute the
498. Service Procedure8.4.2 Fixing the power board • Take the new power board (along with its plate) and fix it (ensure that J3 & J4 connectors
5Table of Contents8.16.1 Removing the Input Switch ...
508. Service Procedure 8.5 Replacing the tube head8.5.1 Removing the tube head• Perform the steps in "8.2.1.1 Power Off:".• Perform the
518. Service Procedure8.5.2 Fixing the new tube head• Take the HV-Tank and fix it to the clamp using the following sequence:-• Using 2.5mm allen ke
528. Service Procedure 8.6 Replacing the Control Board8.6.1 Removing the Control Board• Execute the Steps in “8.2.1.1 Power Off:”.• Execute the St
538. Service Procedure 8.7 Replacing the Control Console8.7.1 Removing the console assembly• Execute the Steps in “8.2.1.1 Power Off:” above.• Exe
54 8.10 Replacing the Scissor Arm cable harness8.10.1 Removing the Scissor arm cables• Execute the Steps in "8.2.1.1 Power Off:".• Execu
558.10.2 Assembly Steps for Scissor-Arm Cable• Hold the end of the Cable harness (Figure 94 ) and insert the communication wire (cable with connec-t
56• Keeping the cable length (towards the L-Clamp end) fixed, route the communication cable near the rotating bush (inside the outer hole of L Arm) a
57• Keeping the vertical arm in vertical position and horizontal arm in horizontal position, route the cable inside the horizontal arm such a way tha
588.10.3 Fixing the Scissor-Arm End caps• For fixing back the scissor-arm end caps first, align the rubber part of scissor-arm as shown in Figure 11
59 8.13 Replacement of Keypad console:8.13.1 Removing the Keypad console:• Execute the steps in “8.2.1.1 Power Off:”.• Pull out the console assemb
6Introduction 1 Introduction 1.1 About Intraskan DCThe Intraskan DC High Frequency Intraoral X-ray has been engineered and manufactured to provide m
60 8.14 Replacement of Doorbell switch:8.14.1 Removing the Door bell switch:• Execute the steps in “8.2.1.1 Power Off:”.• Pull out the Dummy Logo
61 8.16 Replacing the Input switch 8.16.1 Removing the Input Switch• Execute the Steps in "8.2.1.1 Power Off:".• Execute the steps in &q
62• Connect the “Live(Blue) and Neutral(Black) wires(Ring terminals) of switch” to the terminal block fixed on the power board clamp using phillips s
638.17.2 Fixing the new Input power cord• Take the new power cord which comes along with grommet assembly and separate the grommet as-sembly into th
64• Disconnect the console cable from the console board.• Fix back the new console cable on the console board.• Route the cable through the slot pr
65• Re-calibrate the Tube Head by performing the following procedure:• Cover X-ray outlet by lead cap.• Place the Power Switch to the ON position a
66• Switch OFF the power and wait till the Control Console is turned off. Remove the Dead Man Switch Connector Jig which is connected at the bottom o
67 10 Spare Parts ListS. No. Parts Order Code (with Packing Box)1 Tube Head SK3050004602 Tube Head Cover(Top and Bottom) SK3050004613 Control board
68© ImageWorks, Inc Copyright 2013 • Rev. A - October 2013
7Safety and Precautions Installation of the Intraskan DC must be done only by an authorized service engineer. Consult the factory or dealer as neces
82 Safety and Precautions 2.3 Labels on ProductRefer to User Manual for the list and location of each label used on the equipment. 2.2 Safety Symbol
9 3 Product Overview 3.1 Intraskan DC System ComponentsTable 3.1. Intraskan DC System ComponentsDescriptionModelsAvailableIntraskan DC, FS05 , Wall